User’s Manual61XX SeriesPPPPPortion Control Scalesortion Control Scalesortion Control Scalesortion Control Scalesortion Control ScalesModel 6103Model
4. Remove the item from the scale and thedisplayed weight should return to “0.00” andthe scale is ready for the next item.Net Weighing (with a contain
4. When the required weight measurement isachieved, press the ZERO key and the scalewill again display zero “0.00.”5. Continue the same sequence until
The TEST key menu allows access to the followingfeatures: UNDO O, SENSOR (if equipped-SeeSENSOR feature in the CONF Menu Mode),TEST,DONE.When in nor
Press the TEST key to return to normal operationwithout performing any of the above functions.The TEST key is also used to select an item orfeature wh
The 61xx scale powers up in normal weighing modeready for weighing operations. You can access theMENU mode by pressing SW-1 accessed fromthe bottom of
Figure 3Menu Structure12
The Diagnostic (DIAG) Mode menu lets you testspecific areas of the unit’s function.The items in the Diagnostics menu are as follows:Display (DISP) – S
Press the ZERO key to scroll to the I/O test. “I/O” isdisplayed. With a loopback connector in place,press the TEST key to perform the I/O test. “PASS”
From normal weighing mode press SW-1…“DIAG” is displayed.Press the ZERO key until…“CONF” is displayed.Press TEST key to access the Configuration m
From “FILT” press the ZERO key twice…“SENSOR” is displayed (only on equipped models).From this routine you can adjust the sensitivity, viewthe current
UNITED STATESThis equipment has been tested and found to comply with thelimits for a Class A digital device, pursuant to Part 15 of the FCCRules. Thes
To access the Sensor On-Off menu, press the TESTkey. The current state “ON” (or “OFF”) of the photoeyewill be displayed. Pressing the ZERO key willtog
1. From normal weighing mode press SW-1…“DIAG” is displayed. Press the ZERO keyuntil “CAL” is displayed..2. Press the TEST key to access the Calib
6. Place chosen span calibration weight on thescale and wait for the scale to stabilize;then press the TEST key. The display will flashdashes and the
Press the TEST key to access the Units menu. Ifthe scale is currently in the gram unit of measure,the display will flash “CAN’T” and then return todis
*Note: All Jmp1 pins 1, 4, and 6 and all Jmp2pins 7 and 8 are connected internally inside thescale connector.**External Zero Control Line: The optiona
Err-4 Zero ErrorIf a zero error “Err-4” occurs, it is becausethe weight was not stable, or it was undercapacity (lower dashes), or over capacity(up
3. DO NOT submerge the scale base or display inwater.4. DO NOT use acetone or other solvents to cleanthe scale.5. DO NOT expose to extreme temperature
Missing or extra segments on display1. Replace the display board.2. Replace the main board.Scale will not return to zero, or incorrect weight isdispla
Scale is not transmitting data to the host device1. Check cable connection at both the rear of thescale and the host device.2. Check communication set
Display Brackets 6112/6115 Rear Flat 1067-16866 6112 Front/6103 Side Standard 1067-17061 6103 Front 6" Shelf 1067-17064 6112 Front/6103 Side Shel
Table of ContentsDescription... 1Specifications ... 1Initial Se
16" Cable w/Top Exit 7516-xxxxx-366112/6115 Rear Flat Bracket9" Cable w/Top Exit 7516-xxxxx-396103 Side Shelf BracketRemote Zero Footswitch
Notes
1000 Armstrong DriveFairmont, MN 56031Telephone: 507-238-4461Facsimile: 507-238-4195E-mail: [email protected]: www.wt-nci.com
Capacity andResolutionNOTE:NOTE:NOTE:NOTE:NOTE: The number afterthe dash indicates thescale rated capacity in kg.Agency ApprovalsAccuracyDimensionsPow
Input: 120 VAC +10%, -15% Standard 3-wirew/ground inputOptional: 230 VAC +10%, -15% UniversaladapterOutput: 15 VDC @ 0.3 Amps DC minimumUL and CE appr
Initial zero value established on power up if within25% of full scale and the calibrated zero value.Center-zero indicated up to 1/2 division. Fullcapa
1. Check carton for any obvious evidence ofdamage.2. Remove contents from the shipping carton.3. Inspect the unit for shipping damage. Immedi-ately re
8. Verify that the power cable (round end) isplugged into the scale power input connector,then plug the line cord adapter into an appro-priate, proper
On some models, the scale comes equipped withtwo magnetic strips to mount the display to anysuitable surface. Remove the backing from one ofthe magnet