Salter-brecknell HS-300 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Échelles Salter-brecknell HS-300. Salter Brecknell HS-300 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 18
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
HS300
Operation Manual
Contents subject to change without notice
Version 2.1
2012-01-10
Vue de la page 0
1 2 3 4 5 6 ... 17 18

Résumé du contenu

Page 1 - Operation Manual

HS300 Operation Manual Contents subject to change without notice Version 2.1 2012-01-10

Page 2 - EMC compliance

14 Calibration division value Display division value in different weight unit that can be used kg lb Lb:oz (oz) 0.0001kg 0.0001kg 0.0002lb Not a

Page 3 - CONTENT

16 x = 5 : Power-on zero-point ±10%FS; x = 6 : Power-on zero-point ±20%FS; x = 7 : No limitation x = 8 : Power-on zero-point +1%FS x = 9 : Power-

Page 4 - 2. DISPLAY SYMBOL MEANING:

18 Ⅵ. The detail about RS232: 1. RS-232 connects between scale and Host: Scale -----------------------Cable---------------------------Host (DB9 fem

Page 5 - FLT EXT

20 (4) Command: T<CR> (54h 0dh) Response: TARE function is activated, and then returns scale status. just like pressing TARE button: <LF&

Page 6 - CALIBRATION:

22 Table7: (Capacity unit is kg or lb) Resolution set by P7 Division set by P8 and P9(P8=1) 0.0002 0.002 0.02 0.2 2 20 500 0.1000 1.000 10.00 10

Page 7

24 Ⅷ. The meaning of some displayed symbols: 1. 0¯ ¯ ¯ ¯ ---------- zero point is over the setting range 2. 0_ _ _ _ ---------- zero is below the set

Page 8 - . WORKING PARAMETER SETUP:

26 ON/OFF/ZERO + UNIT Normal weighing mode (more than 3s) Enter setup mode when sealed calibration switch is on? ON/OFF/ZERO + TARE Normal weig

Page 9

28 11=Power-on zero-point +4%FS 12=Power-on zero-point +5%FS 13=Power-on zero-point +10%FS 14=Power-on zero-point +20%FS P14.x 0, 1, 2 2 Weight si

Page 10 - Table4: Lb calibration unit:

30 UK and Europe Brecknell Foundry Lane Smethwick West Midlands B66 2LP Te l: +44 (0) 845 246 6717 Fa x : +44 (0) 845 246 6718 Emai

Page 11

Declarations of compliance United States This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant

Page 12 - H1H2H3<CR><ETX>

CONTENT Ⅰ. SPECIFICATIONS. ……………………………………………… P1 Ⅱ. KEY FUNCTION. ……………………………………………… P3 Ⅲ. CALIBRATION. ………………………………….……………… P6 Ⅳ. Examination

Page 13

2 ·OPERATION INTERFACE: 1. Display: 0.65” (17mm) 7-segment LCD, 51/2 digits 2. Keyboard: 4-key push button ·POWER: 1. Alkaline Batteries: 4 x “AAA

Page 14

4 (2) Choose weighing units among kg-lb-lb: oz Note: The weighing units that can be used are restricted by display division, and calib

Page 15

6 (2) Choose the weight inner code or input working voltage to be displayed. The working voltage range is 4.8V-8V. If the voltage is not within ran

Page 16

8 can press ON/OFF/ZERO to exit the calibration mode. Or load 5%-100%FS weight, and use PRINT/HOLD and UNIT keys to input the loaded weight. If the i

Page 17

10 Below is the drawing of position of the zero-point potentiometer on PCB for conference, decrease ADC data by rotating clockwise, and increase ADC

Page 18

12 as follows: <LF><Gross: reading, minus, decimal point, unit><CR><EXT> <LF> <Tare: reading, decimal point, unit&

Commentaires sur ces manuels

Pas de commentaire